Sonoclot (SC1) ACT Machine incorporates Sonoclot's Proprietary Visco-Elastic technology and is extremely sensitive to the developing clot.
Normal ACT Machines mechanically detect mature clot to report Activated Clotting Time which takes into account cumulative activity of Factors, Fibrin gel & Platelets to form a mature clot.
On the other hand Sonoclot ACT Machine starts monitoring the viscosity changes, which is the first component to change when the coagulation cascade is triggered by the respective activators.
Factor Xa triggers Prothrombin to Thrombin (Factor IIa) the sample is still in the liquid phase. As Thrombin forms it triggers Fibrinogen to Fibrin Polymerization the sample starts to gel and the viscosity starts increasing. Sonoclot ACT Machine is very sensitive to this changing viscosity and first reports ACT, which is a function of Factor Xa activity.
Once the ACT results are obtained, Sonoclot ACT Machine measures the Gel formation phase (Fibrinogen to Fibrin Polymerization) and reports the same as Clot Rate. Clot Rate (Rate of Actual Clot Formation) is rate of Polymerization of Fibrinogen to Fibrin and is an indicator of activity of Fibrinogen, Factor IIa (Thrombin) generation, and Thrombin Inhibitors such as ATIII. This result has particular importance in evaluating the Thrombin Inhibition & Heparin Anti-Coagulant Activity. Clot Rate is an indicator of Level of Thrombin Inhibition which is very useful in cases of Heparin Resistance, Low or High Thrombin Inhibition, Stent Thrombosis EVEN WHEN ACT IS SAME.